@article{oai:asahi-u.repo.nii.ac.jp:00012201, author = {ADACHI, KOTONO and Yamamoto, Hiroaki and 渡邊, 諒 and 岩堀, 正俊 and IWAHORI, MASATOSHI and 都尾, 元宣 and MIYAO, MOTONOBU and 山内, 六男 and YAMAUCHI, MUTSUO}, issue = {2020-02}, journal = {2020-02}, month = {2020-02, 2020-03-30}, note = {We examined the reproducibility of measurement values obtained using the recently released Dental Prescale II system (50H, GC) and analysis software (Bite Force Analyzer, GC). One adult male volunteer with normal occlusion was asked to bite down for 3 s with the Prescale II in situ to produce the sample. Occlusal contact area, mean occlusal pressure, and occlusal force were measured. The measurements were then analyzed 5 times each under the following conditions: 1) with the sample placed on the scanner (reference condition); 2) after removing and then replacing the sample on the scanner (repositioned); 3) after automated trimming (adjusted); 4) after manual trimming on the screen (manually adjusted); and 5) after the sample was stored for 2 days indoors (day 2) or 3 days indoors (day 3). Statistical analysis was performed using one-way analysis of variance with Bonferroni post hoc analysis. The coefficient of variation (CV) was also determined. Measurement values for occlusal contact area were significantly lower under the reference condition than under the repositioned, adjusted, manually adjusted, and day 3 conditions. Mean occlusal pressure was significantly higher under all conditions compared with the reference condition. Occlusal force was significantly lower under the repositioned, adjusted, and manually adjusted conditions compared with the reference condition but was significantly higher under the day 2 condition. CV values were 0.71-4.48 for occlusal contact area, 0.46-1.23 for mean occlusal pressure, and 0.28-1.05 for occlusal force. These results indicate high reliability of the Dental Prescale II system and occlusal force analysis software.}, pages = {2020-02--2020-02}, title = {Reproducibility of the Newly Developed Dental Prescale II System and Bite Force Analyzer for Occlusal Measurements}, volume = {2020-02}, year = {}, yomi = {WATANABE, RYO} }